Quality control solution for the nova lactate meter.

Run 2 different levels of the StatStrip Lactate Control Solutions during each 24 hours of testing prior to testing of patient specimens and under the following circumstances:

  • If a patient test has been repeated and the blood lactate results are lower or higher than expected
  • If there are other indications that the system is not working properly 30
  • As required by the institution’s quality control policy or local regulatory requirements
  • Anytime there is a concern the accuracy of the meter may have been affected by rough handling (such as dropping the meter)

Gebrauchsanweisung

How to perform quality control?

  1. Insert a test strip into the meter. All segments of the screen will display for 2 seconds. Then a flashing blood drop will display.
  2. Identify the sample as a Control; use the Left or Right button to find the desired control level: C1 and C2.
  3. Touch the end of the test strip to a drop of control solution until the test strip fills and the meter beeps.
  4. Lactate quality control test results are available on screen in 13 seconds
  5. There is one long beep when the results are ready. There are 3 short beeps if test results are outside the measurement range of the test strip.